1530 papers on microsampling and monitoring, each with a finding paraphrased to our standard and labelled in one vocabulary. Every entry links to a legitimate copy; nothing is copied from an abstract.

12 papers labelled “yourbio”, newest first.

  1. 2026

    In 480 adults, capillary specimens from fingertip lancets or the TAP Micro Select device gave clinically comparable results to venous samples for 16 of 18 routine chemistry analytes; potassium and chloride were not interchangeable, and fingertip collection showed more haemolysis, both points to weigh before decentralised microsampling of routine chemistry.

    Venous-capillary blood agreement for routine chemistry analytes: effects of collection site and sample matrixDeza et al., Clinical chemistry and laboratory medicine (paywalled)

    • venous-agreement
    • blood
    • capillary
    • yourbio
    • validation
  2. 2025

    A head-to-head of four upper-arm capillary devices, including TAP Micro Select, versus fingerstick and venous: experience and performance varied by device with none clearly superior, while analyte correlations with venous were similar across all.

    Comparing capillary blood collection technologies: patient experience, device performance and clinical accuracyCollier et al., Bioanalysis (paywalled)

    • blood
    • liquid
    • yourbio
    • capillary
    • venous-agreement
    • acceptability
    • validation
    • bd-microtainer
  3. 2025

    In 46 women under hormone monitoring at Columbia University Fertility Center, oestradiol, LH, FSH, progesterone and hCG measured on routine immunoassays from 400 to 600 microlitres of upper-arm capillary blood collected with TAP Micro Select agreed closely with paired venous results for every analyte, with correlation and concordance high throughout. It is the evidence that puts a self-collected sample inside the daily monitoring of an IVF cycle.

    Painless, needle-free capillary blood collection for fertility hormone monitoringLattin et al., Fertility and sterility (paywalled)

    • blood
    • liquid
    • capillary
    • yourbio
    • fertility
    • hormones
    • venous-agreement
    • validation
  4. 2024

    In 91 children and adults, capillary C-peptide taken with the TAP device matched venous sampling with 100% sensitivity and specificity for a clinically relevant level, 200 pmol/L or above, with r = 0.996. 63% preferred it to a needle against 7% who preferred the needle, and same-day bruising was 5.5% versus 34.8% after a venous draw.

    Transdermal blood sampling for C-peptide is a minimally invasive, reliable alternative to venous sampling in children and adults with type 1 diabetesBesser et al., Diabetes Care

    • blood
    • liquid
    • yourbio
    • capillary
    • pediatric
    • venous-agreement
    • acceptability
    • validation
  5. 2024

    Capillary sampling gave TSH and FT4 results that closely matched venous values, with Spearman correlations of 0.99 and 0.97 respectively. This supports using small capillary samples for remote thyroid function testing.

    Determination of Capillary Blood TSH and Free Thyroxine Levels Using Digital ImmunoassaySuzuki et al., Journal of the Endocrine Society

    • venous-agreement
    • blood
    • capillary
    • hormones
    • yourbio
  6. 2024

    Capillary sampling for carcinoembryonic antigen showed near perfect agreement with venous results and was preferred by most participants, supporting reliable home based collection that could reduce hospital visits.

    Feasibility, reliability and satisfaction of (automated) capillary carcinoembryonic antigen measurements for future home-based blood sampling: the prospective CASA-I studyVoigt et al., Colorectal disease : the official journal of the Association of Coloproctology of Great Britain and Ireland (paywalled)

    • venous-agreement
    • acceptability
    • blood
    • self-collection
    • oncology
    • colorectal
    • capillary
    • yourbio
  7. 2023

    TAP II capillary serum, both professionally collected and self-collected, correlated with venous serum at R > 0.9 for eight analytes including ALT, AST, cholesterol, HDL and triglycerides. Creatinine correlated acceptably but carried a consistent negative bias, and carbon dioxide, potassium and albumin did not reach acceptable agreement: the useful reading is that the panel has to be validated, not the device alone.

    Maximizing microsampling: measurement of comprehensive metabolic and lipid panels using a novel capillary blood collection deviceCollier et al., The Journal of Applied Laboratory Medicine (paywalled)

    • venous-agreement
    • blood
    • self-collection
    • capillary
    • bd-microtainer
    • liquid
    • yourbio
    • validation
    • biomarkers
  8. 2022

    The paper records that the TAP II devices used for the infant draws were provided by YourBio Health, and that capillary serum was collected from babies as young as two months: the smallest patients sampled, without a heel prick or a venepuncture.

    Durability of anti-spike antibodies in infants after maternal COVID-19 vaccination or natural infectionShook, Edlow et al., JAMA

    • blood
    • liquid
    • yourbio
    • capillary
    • pediatric
    • serology
  9. 2022

    Patients self-collected upper-arm capillary blood with press-activated devices including TAP II; autoantibody and CRP results agreed with venous at r≥0.98, were rated less painful, and 98.6% got enough blood within two attempts.

    Remote self-collection of upper-arm capillary blood for autoantibody analysis in immune-mediated rheumatic diseasesZarbl et al., RMD Open

    • venous-agreement
    • acceptability
    • blood
    • self-collection
    • capillary
    • liquid
    • yourbio
    • biomarkers
  10. 2022

    The TAP device showed near perfect agreement with venous sampling for anti-Müllerian hormone measurements, with higher accuracy and lower pain compared to ADx card. This validates TAP as a reliable option for patient-led fertility hormone testing at home.

    Evaluation of at-home serum anti-Müllerian hormone testing: a head-to-head comparison studySilliman et al., Reproductive biology and endocrinology : RB&E

    • blood
    • capillary
    • yourbio
    • venous-agreement
    • acceptability
    • fertility
    • hormones
  11. 2020

    Protein abundances measured by SomaScan were comparable between capillary samples collected with the TAP device and venous samples, and selected ELISA assays showed strong correlation, supporting the use of the TAP device for quantitative protein biomarker analysis in decentralised trials.

    Evaluation of a novel blood microsampling device for clinical trial sample collection and protein biomarker analysisXing et al., Bioanalysis (paywalled)

    • dct
    • venous-agreement
    • blood
    • capillary
    • yourbio
    • biomarkers
  12. 2018

    Metabolomics of TAP capillary blood matched venous draws for most metabolites, amino acids, bile acids and acyl-carnitines at r≥0.95; a small redox/energy subset differed with collection dynamics.

    Quantitative metabolomics comparison of traditional blood draws and TAP capillary blood collectionCatala et al., Metabolomics (paywalled)

    • venous-agreement
    • blood
    • metabolome
    • capillary
    • liquid
    • yourbio
    • validation