Comparing capillary blood collection technologies: patient experience, device performance and clinical accuracy

Collier et al.

The finding, in our words

A head-to-head of four upper-arm capillary devices, including TAP Micro Select, versus fingerstick and venous: experience and performance varied by device with none clearly superior, while analyte correlations with venous were similar across all.

A paraphrase to the Library’s standard, never the abstract. The source is one link away and is always the authority.

Labels

  1. 2025

    A vacuum-assisted upper-arm device collected a mean 538 µL and met the volume target in 85.5% of cases; 17 of 19 analytes met CLIA agreement with venous, with pain lower than fingerstick.

    Evaluation of a Simplified Upper-Arm Device for Vacuum-Assisted Collection of Capillary BloodSchaff et al., Diagnostics · source ↗

    • blood
    • liquid
    • capillary
    • venous-agreement
    • acceptability
    • validation
    • bd-microtainer
  2. 2024

    In 91 children and adults, capillary C-peptide taken with the TAP device matched venous sampling with 100% sensitivity and specificity for a clinically relevant level, 200 pmol/L or above, with r = 0.996. 63% preferred it to a needle against 7% who preferred the needle, and same-day bruising was 5.5% versus 34.8% after a venous draw.

    Transdermal blood sampling for C-peptide is a minimally invasive, reliable alternative to venous sampling in children and adults with type 1 diabetesBesser et al., Diabetes Care · source ↗

    • blood
    • liquid
    • yourbio
    • capillary
    • pediatric
    • venous-agreement
    • acceptability
    • validation
  3. 2023

    TAP II capillary serum, both professionally collected and self-collected, correlated with venous serum at R > 0.9 for eight analytes including ALT, AST, cholesterol, HDL and triglycerides. Creatinine correlated acceptably but carried a consistent negative bias, and carbon dioxide, potassium and albumin did not reach acceptable agreement: the useful reading is that the panel has to be validated, not the device alone.

    Maximizing microsampling: measurement of comprehensive metabolic and lipid panels using a novel capillary blood collection deviceCollier et al., The Journal of Applied Laboratory Medicine (paywalled) · source ↗

    • venous-agreement
    • blood
    • self-collection
    • capillary
    • bd-microtainer
    • liquid
    • yourbio
    • validation
    • biomarkers
  4. 2026

    Although most kidney transplant recipients successfully collected capillary blood samples at home, haemolysis and transport conditions affected analyte stability. Agreement with venous sampling was acceptable for haemoglobin and creatinine, but potassium and haematocrit exceeded total allowable error limits.

    At-home blood collection for clinical chemistry analyses in a kidney transplant population: a feasibility studyHoekstra et al., Clinical chemistry and laboratory medicine (paywalled) · source ↗

    • blood
    • liquid
    • capillary
    • bd-microtainer
    • self-collection
    • validation
    • venous-agreement
  5. 2026

    Capillary blood collected with Tasso+ devices demonstrated strong agreement with venous blood for primary Athlete Biological Passport markers (haemoglobin, reticulocytes, OFF-score, ABPS) with correlations ≥0.9, though platelet count showed poor concordance. Despite 44 of 209 samples failing pre-analytical checks mainly due to coagulation, high athlete acceptance and minimal pain support capillary sampling as a less invasive, viable alternative for decentralised anti-doping controls.

    Capillary Blood as a Potential New Matrix for the Athlete Biological Passport: A Field-Based Comparison StudyRequena-Tutusaus et al., Medicine and science in sports and exercise (paywalled) · source ↗

    • blood
    • liquid
    • capillary
    • validation
    • venous-agreement
    • acceptability
    • doping