2022 · Journal of pharmaceutical and biomedical analysis · paywalled
Comparison of conventional dried blood spots and volumetric absorptive microsampling for tacrolimus and mycophenolic acid determination
Paniagua-González et al.
The finding, in our words
Clinical validation for tacrolimus and mycophenolic acid monitoring showed that while volumetric absorptive microsampling met analytical criteria after concentration correction, conventional dried blood spots achieved superior adherence to strict clinical criteria, sample quality, and cost efficiency without significant haematocrit bias.
A paraphrase to the Library’s standard, never the abstract. The source is one link away and is always the authority.
In AML and CLL, 91% of VAMS venetoclax results fell within 20% of plasma after individualised haematocrit correction; in home sampling, 18 of 21 patients self-sampled independently and 76% of returned samples were analysable, which shows home microsampling is workable, though the authors ask for multicentre validation.
This review found that volumetric microsampling devices, such as Mitra, HemaPEN, HemaXis DB10, and Tasso-M20, overcome haematocrit bias, whilst Telimmune plasma separation cards enable direct plasma collection without centrifugation. These technologies support patient-centric, decentralised sampling, though the authors note that clinical validation remains limited across different drugs and patient populations.
In matched clinical samples from rheumatoid arthritis, VAMS and DBS showed strong agreement for methotrexate polyglutamates (slopes 0.95-1.07; bias within -4.21% to 0.36%; SRCC ≥ 0.969), with up to 100% of total MTXPG results within ±20% limits; capillary microsampling agreed closely with whole blood but differed from red blood cells, indicating matrix-specific differences that must be accounted for when interpreting against RBC-based reference values.
A validated LC-MS/MS method for everolimus using Mitra and Capitainer devices found excellent agreement between capillary microsamples and venous whole blood in 33 adult transplant recipients. The results showed high accuracy and negligible haematocrit effects, supporting the use of microsampling for decentralised therapeutic drug monitoring.
The study validated a VAMS-based LC-MS/MS method for measuring sirolimus levels, showing it was clinically equivalent to the standard whole blood method and suitable for therapeutic drug monitoring in paediatric transplant patients, supporting the use of decentralised microsampling for this purpose.