VAMS microsampling found no significant difference from venous blood for tacrolimus in paediatric patients; samples were stable for 14 days, with average difference between paired at-home samples of 0.12 ± 0.94 ng/mL.
Clinical validation for tacrolimus and mycophenolic acid monitoring showed that while volumetric absorptive microsampling met analytical criteria after concentration correction, conventional dried blood spots achieved superior adherence to strict clinical criteria, sample quality, and cost efficiency without significant haematocrit bias.
In lung and renal transplant recipients, VAMS sampling with LC-MS/MS quantification of mycophenolic acid and tacrolimus showed good linearity and accuracy, and with a haematocrit-adjusted conversion formula achieved clinical agreement in most samples; tacrolimus did not require haematocrit correction. The approach is virtually painless and enables richer sampling for more accurate drug exposure estimates, supporting decentralised therapeutic drug monitoring.
This review found that volumetric microsampling devices, such as Mitra, HemaPEN, HemaXis DB10, and Tasso-M20, overcome haematocrit bias, whilst Telimmune plasma separation cards enable direct plasma collection without centrifugation. These technologies support patient-centric, decentralised sampling, though the authors note that clinical validation remains limited across different drugs and patient populations.
In matched clinical samples from rheumatoid arthritis, VAMS and DBS showed strong agreement for methotrexate polyglutamates (slopes 0.95-1.07; bias within -4.21% to 0.36%; SRCC ≥ 0.969), with up to 100% of total MTXPG results within ±20% limits; capillary microsampling agreed closely with whole blood but differed from red blood cells, indicating matrix-specific differences that must be accounted for when interpreting against RBC-based reference values.