Validation of a Capillary Dry Blood Sample MITRA-Based Assay for the Quantitative Determination of Systemic Tacrolimus Concentrations in Transplant Recipients
Undre et al.
The finding, in our words
Capillary blood collected with the Neoteryx Mitra device showed a high correlation with venous samples for tacrolimus monitoring in transplant recipients, but yielded concentrations that were on average 22.5% higher. This minimally invasive method offers a convenient alternative to venepuncture for therapeutic drug monitoring, provided the consistent positive bias is accounted for in clinical interpretation.
A paraphrase to the Library’s standard, never the abstract. The source is one link away and is always the authority.
In matched clinical samples from rheumatoid arthritis, VAMS and DBS showed strong agreement for methotrexate polyglutamates (slopes 0.95-1.07; bias within -4.21% to 0.36%; SRCC ≥ 0.969), with up to 100% of total MTXPG results within ±20% limits; capillary microsampling agreed closely with whole blood but differed from red blood cells, indicating matrix-specific differences that must be accounted for when interpreting against RBC-based reference values.
Kocur et al., International journal of molecular sciences · source ↗
A validated LC-MS/MS method for everolimus using Mitra and Capitainer devices found excellent agreement between capillary microsamples and venous whole blood in 33 adult transplant recipients. The results showed high accuracy and negligible haematocrit effects, supporting the use of microsampling for decentralised therapeutic drug monitoring.
A validated offline SPE-LC-MS/MS method using the Mitra microsampling device successfully quantified tacrolimus, cyclosporine A, tryptophan, kynurenine, and creatinine simultaneously. The method met EMA and FDA criteria, showing agreement between capillary and venous sampling, which could facilitate decentralised therapeutic drug monitoring and transplant follow-up through patient self-collection.
Nierychlewski et al., Therapeutic drug monitoring · source ↗
The study validated a VAMS-based LC-MS/MS method for measuring sirolimus levels, showing it was clinically equivalent to the standard whole blood method and suitable for therapeutic drug monitoring in paediatric transplant patients, supporting the use of decentralised microsampling for this purpose.
Kocur et al., Pharmacological reports : PR · source ↗
A study of 50 paediatric renal transplant recipients found that a validated VAMS LC-MS/MS method for tacrolimus showed high correlation with the reference method over 0.5–60 ng/mL, with minimal haematocrit effect.