2025 · Therapeutic drug monitoring · open access
An Offline SPE-LC-MS/MS Method for Simultaneous Quantification of Tacrolimus, Cyclosporine A, Kynurenine, Tryptophan, and Creatinine Using Volumetric Absorptive Microsampling Device Mitra
Nierychlewski et al.
The finding, in our words
A validated offline SPE-LC-MS/MS method using the Mitra microsampling device successfully quantified tacrolimus, cyclosporine A, tryptophan, kynurenine, and creatinine simultaneously. The method met EMA and FDA criteria, showing agreement between capillary and venous sampling, which could facilitate decentralised therapeutic drug monitoring and transplant follow-up through patient self-collection.
A paraphrase to the Library’s standard, never the abstract. The source is one link away and is always the authority.