2019 · Journal of chromatography. B, Analytical technologies in the biomedical and life sciences · paywalled
Validation and clinical application of an LC-MS/MS method for the quantification of everolimus using volumetric absorptive microsampling
Verheijen et al.
The finding, in our words
The VAMS method for everolimus quantification met validation criteria and samples remained stable for a year at room temperature, but bias of -20 to 31% across 30-50% haematocrit ranges meant the expected advantage over dried blood spots was not demonstrated. However, using an empirically derived formula, whole blood concentrations could be back-calculated with reasonable accuracy, supporting its potential for decentralised therapeutic drug monitoring.