Self-collection of capillary blood using Tasso-SST devices for Anti-SARS-CoV-2 IgG antibody testing

Hendelman et al.

The finding, in our words

Capillary blood self‑collection with the Tasso‑SST device succeeded in over 93 % of attempts and produced SARS‑CoV‑2 IgG results that agreed closely with venous blood, maintaining 98.9 % overall accuracy after simulated summer and winter shipping. This supports reliable, decentralised serology testing without medical supervision.

A paraphrase to the Library’s standard, never the abstract. The source is one link away and is always the authority.

Labels

  1. 2026

    Capillary blood samples collected by staff or self-collected by participants showed strong correlation and clinically acceptable agreement with venipuncture for influenza haemagglutination inhibition titers. The small differences observed were not clinically meaningful, and adequate sample volume was achieved in most attempts, supporting the feasibility of decentralised serological testing.

    Comparison of influenza hemagglutination inhibition titers from capillary blood versus venipunctureRick et al., Journal of clinical virology : the official publication of the Pan American Society for Clinical Virology (paywalled) · source ↗

    • blood
    • liquid
    • capillary
    • self-collection
    • validation
    • venous-agreement
    • serology
  2. 2026

    In a cohort of 62 participants evaluating home self-collection with the TassoPlus device, 38 returned samples, with 29% excluded due to insufficient plasma volume below 200 ΔL. Although viral load measurements in adequate samples correlated well with conventional testing (r = 0.800), the high failure rate and missed detection in low-volume viremic samples indicate that further technical refinements are necessary before clinical adoption.

    Feasibility and reliability of decentralized HIV-1 viral load monitoring on self-sampled bloodMendes de Leon et al., PloS one · source ↗

    • blood
    • liquid
    • capillary
    • self-collection
    • validation
    • venous-agreement
    • acceptability
    • serology
  3. 2025

    A study of decentralised patient-centric microsampling found that 82% of untrained participants, 47 of 57, successfully self-collected acceptable fingerstick blood in Microtainer tubes. All 46 tested samples yielded accurate HIV viral suppression results compared to venipuncture, though this remote monitoring approach is limited to US courier services and temperature conditions.

    A cross-sectional analysis of fingerstick blood self-microcollection for remote HIV suppression monitoring in Atlanta, Georgia, USA: a path to expanding access to continuum of careJohnson et al., AIDS research and therapy · source ↗

    • blood
    • liquid
    • capillary
    • bd-microtainer
    • self-collection
    • validation
    • venous-agreement
    • serology
  4. 2024

    In 30 adult solid organ transplant recipients, capillary blood self-collected using the Tasso+ device showed 91% concordance with venous samples for detecting CMV DNAemia above the limit of detection, with excellent quantitative correlation (Spearman R² = 0.99). Most participants preferred self-collection over venipuncture, suggesting the device could enable decentralised viral monitoring, though sensitivity was lower (95% at 308 IU/mL).

    Evaluation of the Tasso+ blood self-collection device for quantitation of plasma cytomegalovirus (CMV) DNAemia in adult solid organ transplant recipients (SOTr)Phan et al., Microbiology spectrum · source ↗

    • blood
    • liquid
    • capillary
    • self-collection
    • validation
    • venous-agreement
    • acceptability
    • serology
  5. 2022

    Capillary blood self-sampling achieved 98.3% agreement with venous samples for SARS-CoV-2 IgG antibodies and 100% for IgA among 60 participants, demonstrating accuracy for vaccine monitoring. Saliva self-collection proved feasible with excellent usability, offering a patient-centric alternative for decentralised immunogenicity assessment.

    Self-collection of capillary blood and saliva to determine COVID-19 vaccine immunogenicity in patients with immune-mediated inflammatory diseases and health professionalsSchmetzer et al., Frontiers in public health · source ↗

    • blood
    • saliva
    • liquid
    • capillary
    • self-collection
    • validation
    • venous-agreement
    • acceptability
    • serology