"Center punch" and "whole spot" bioanalysis of apixaban in human dried blood spot samples by UHPLC-MS/MS
Zheng et al.
The finding, in our words
Validation of two dried blood spot methods for apixaban showed that whole-spot analysis eliminated haematocrit and lipid interference while maintaining acceptable accuracy and precision. This robustness supports decentralised therapeutic drug monitoring from capillary finger-prick samples.
A paraphrase to the Library’s standard, never the abstract. The source is one link away and is always the authority.
In AML and CLL, 91% of VAMS venetoclax results fell within 20% of plasma after individualised haematocrit correction; in home sampling, 18 of 21 patients self-sampled independently and 76% of returned samples were analysable, which shows home microsampling is workable, though the authors ask for multicentre validation.
Levens et al., Clinical Pharmacokinetics (paywalled) · source ↗
This review found that volumetric microsampling devices, such as Mitra, HemaPEN, HemaXis DB10, and Tasso-M20, overcome haematocrit bias, whilst Telimmune plasma separation cards enable direct plasma collection without centrifugation. These technologies support patient-centric, decentralised sampling, though the authors note that clinical validation remains limited across different drugs and patient populations.
Huhn & Scherf-Clavel, Therapeutic drug monitoring (paywalled) · source ↗
In matched clinical samples from rheumatoid arthritis, VAMS and DBS showed strong agreement for methotrexate polyglutamates (slopes 0.95-1.07; bias within -4.21% to 0.36%; SRCC ≥ 0.969), with up to 100% of total MTXPG results within ±20% limits; capillary microsampling agreed closely with whole blood but differed from red blood cells, indicating matrix-specific differences that must be accounted for when interpreting against RBC-based reference values.
Kocur et al., International journal of molecular sciences · source ↗
The study found a dried blood spot method for ustekinumab was linear from 3 to 12 mg/L with inter-day accuracy of 90.1 to 106%, enabling reliable capillary blood measurement for remote monitoring; the authors note the single-centre design.
The HematoCARD cartridge autonomously collects duplicate 10 microlitre dried blood spot samples from a single finger prick and shows good analytical performance for monitoring adalimumab, with accuracy between 91 and 111 per cent, linearity R squared 0.99 and coefficient of variation at or above 0.94 compared with serum reference and commercial DBS microsampling methods.
Van Hileghem et al., Analytical chemistry (paywalled) · source ↗