Application of volumetric absorptive microsampling device for quantification of tacrolimus in human blood as a model drug of high blood cell partition
Kita & Mano
The finding, in our words
Volumetric absorptive microsampling was qualified to quantify tacrolimus across a clinically relevant range with acceptable linearity, precision and extraction recovery despite haematocrit variation, but samples should be stored at -25 degrees Celsius to avoid loss of recovery at ambient or refrigerated temperatures.
A paraphrase to the Library’s standard, never the abstract. The source is one link away and is always the authority.
In lung and renal transplant recipients, VAMS sampling with LC-MS/MS quantification of mycophenolic acid and tacrolimus showed good linearity and accuracy, and with a haematocrit-adjusted conversion formula achieved clinical agreement in most samples; tacrolimus did not require haematocrit correction. The approach is virtually painless and enables richer sampling for more accurate drug exposure estimates, supporting decentralised therapeutic drug monitoring.
Juan et al., Journal of pharmacy & pharmaceutical sciences : a publication of the Canadian Society for Pharmaceutical Sciences, Societe canadienne des sciences pharmaceutiques · source ↗
In matched clinical samples from rheumatoid arthritis, VAMS and DBS showed strong agreement for methotrexate polyglutamates (slopes 0.95-1.07; bias within -4.21% to 0.36%; SRCC ≥ 0.969), with up to 100% of total MTXPG results within ±20% limits; capillary microsampling agreed closely with whole blood but differed from red blood cells, indicating matrix-specific differences that must be accounted for when interpreting against RBC-based reference values.
Kocur et al., International journal of molecular sciences · source ↗
A validated LC-MS/MS method for everolimus using Mitra and Capitainer devices found excellent agreement between capillary microsamples and venous whole blood in 33 adult transplant recipients. The results showed high accuracy and negligible haematocrit effects, supporting the use of microsampling for decentralised therapeutic drug monitoring.
The study validated a VAMS-based LC-MS/MS method for measuring sirolimus levels, showing it was clinically equivalent to the standard whole blood method and suitable for therapeutic drug monitoring in paediatric transplant patients, supporting the use of decentralised microsampling for this purpose.
Kocur et al., Pharmacological reports : PR · source ↗
A study of 50 paediatric renal transplant recipients found that a validated VAMS LC-MS/MS method for tacrolimus showed high correlation with the reference method over 0.5–60 ng/mL, with minimal haematocrit effect.