Using matrix-induced ion suppression combined with LC-MS/MS for quantification of trimethylamine-N-oxide, choline, carnitine and acetylcarnitine in dried blood spot samples
Lu et al.
The finding, in our words
A dried-blood-spot method using matrix-induced ion suppression was validated to estimate sample volume and correct for haematocrit effects, achieving 91-113% quantification accuracy for TMAO, choline, carnitine and acetylcarnitine. This enables reliable, decentralised testing of gut-microbiome-derived metabolites linked to cardiovascular disease.
A paraphrase to the Library’s standard, never the abstract. The source is one link away and is always the authority.
Microsampling enables less invasive, patient-centric self-collection of capillary blood for remote monitoring of metabolites and lipids, overcoming conventional venipuncture constraints. Recent device innovations address dried blood spot limitations, particularly haematocrit and volume variations, expanding decentralised applications in population health, drug discovery and multi-omics research.
Volumetric absorptive microsampling showed better precision than dried blood spots and a metabolic profile closer to whole blood, with stable signalling lipids for 24 hours at room temperature but significant changes after one week unless stabilised, indicating feasibility for decentralised sampling with the need for storage strategies.
Thangavelu et al., Analytical and bioanalytical chemistry · source ↗
Capillary dried blood spots, after haematocrit-dependent conversion, showed good agreement with plasma for 25-hydroxyvitamin D quantification, with 90 per cent of results within 20 per cent of plasma and substantial to almost perfect agreement in status classification, supporting reliable home self-collection for large-scale vitamin D monitoring.
Heughebaert et al., Clinical chemistry and laboratory medicine (paywalled) · source ↗
Capillary finger-stick dried blood spots demonstrated strong analytical agreement with venous serum for prostate-specific antigen (R² = 0.987) and remained stable for 31 days across a wide temperature range. This less invasive microsampling approach enables at-home self-collection, supporting decentralised screening and tele-diagnostics for prostate cancer.
van Vugt et al., The journal of applied laboratory medicine (paywalled) · source ↗
Untargeted metabolomic profiles from three blood microsampling devices aligned more closely with whole blood than with plasma, and all devices distinguished sex based on amino acids, lipids, and acylcarnitines. This validates that device choice can be tailored to the metabolites of interest for decentralised human biomonitoring.
Avella et al., Metabolomics : Official journal of the Metabolomic Society · source ↗