2025 · Journal of general and family medicine · open access
Comparison between point-of-care testing from capillary samples and conventional laboratory testing from venous samples for white blood cells and C-reactive protein in a pediatric outpatient setting
Kuniyoshi et al.
The finding, in our words
In 277 paediatric patients, capillary microsampling with the Microsemi CRP device found close agreement with venous testing for CRP, with a mean difference of -0.25 mg/dL and 95% limits of -2.1 to 1.6 mg/dL. WBC showed a bias of -18 × 100/μL, with 95% limits of -73 to 37 × 100/μL. This supports decentralised, patient-centric screening of febrile children.
A paraphrase to the Library’s standard, never the abstract. The source is one link away and is always the authority.