An at-home blood collection device for remote immune monitoring by high-parameter flow cytometry
Konecny et al.
The finding, in our words
Capillary blood obtained using an at-home collection device matched venous blood closely in immune cell composition and phenotype. Although shipping affected myeloid cells, T cell numbers, subsets, and phenotypes stayed stable, which supports the practicality of remote immune monitoring.
A paraphrase to the Library’s standard, never the abstract. The source is one link away and is always the authority.
This UK consensus opinion outlines how clinical laboratories can validate self-collected capillary blood sampling against venous reference standards to achieve accreditation and facilitate the shift towards community-based patient care.
Woolley et al., Annals of clinical biochemistry (paywalled) · source ↗
Capillary blood samples collected by staff or self-collected by participants showed strong correlation and clinically acceptable agreement with venipuncture for influenza haemagglutination inhibition titers. The small differences observed were not clinically meaningful, and adequate sample volume was achieved in most attempts, supporting the feasibility of decentralised serological testing.
Rick et al., Journal of clinical virology : the official publication of the Pan American Society for Clinical Virology (paywalled) · source ↗
Self-collected capillary blood samples posted by standard mail showed high concordance with couriered venous samples for HbA1c, creatinine and lipid parameters, supporting their use for remote monitoring in diabetes management and screening.
Hoffmeister et al., Journal of diabetes investigation · source ↗
Although most kidney transplant recipients successfully collected capillary blood samples at home, haemolysis and transport conditions affected analyte stability. Agreement with venous sampling was acceptable for haemoglobin and creatinine, but potassium and haematocrit exceeded total allowable error limits.
Hoekstra et al., Clinical chemistry and laboratory medicine (paywalled) · source ↗
In a cohort of 62 participants evaluating home self-collection with the TassoPlus device, 38 returned samples, with 29% excluded due to insufficient plasma volume below 200 ΔL. Although viral load measurements in adequate samples correlated well with conventional testing (r = 0.800), the high failure rate and missed detection in low-volume viremic samples indicate that further technical refinements are necessary before clinical adoption.