Minimally invasive capillary blood sampling methods

Hoffman et al.

The finding, in our words

This review evaluates established and emerging capillary blood sampling techniques, comparing factors such as wound geometry and patient comfort, and identifies microneedles as a promising option for paediatric care because they can obtain samples virtually painlessly.

A paraphrase to the Library’s standard, never the abstract. The source is one link away and is always the authority.

Labels

  1. 2025

    Dried blood spot sampling matched venous serum performance for islet autoantibody detection and was considered minimally invasive and convenient by parents and stakeholders, supporting its use for decentralised screening in home or community settings.

    Establishing the performance and acceptability of dried blood spot sampling to screen for islet-specific autoantibodiesFaustini et al., Diabetic medicine : a journal of the British Diabetic Association · source ↗

    • blood
    • dried
    • capillary
    • dbs
    • self-collection
    • validation
    • venous-agreement
    • acceptability
    • pediatric
  2. 2024

    In 91 children and adults, capillary C-peptide taken with the TAP device matched venous sampling with 100% sensitivity and specificity for a clinically relevant level, 200 pmol/L or above, with r = 0.996. 63% preferred it to a needle against 7% who preferred the needle, and same-day bruising was 5.5% versus 34.8% after a venous draw.

    Transdermal blood sampling for C-peptide is a minimally invasive, reliable alternative to venous sampling in children and adults with type 1 diabetesBesser et al., Diabetes Care · source ↗

    • blood
    • liquid
    • yourbio
    • capillary
    • pediatric
    • venous-agreement
    • acceptability
    • validation
  3. 2024

    In at-home and trial settings, a wet capillary device reliably collected a median ~450 µL with a 4.4% failure rate; 88% of adults preferred self-collection over venipuncture, with lower pain.

    Feasibility, acceptability and safety of a device for self-collecting capillary blood in clinical trialsDasari et al., PLOS ONE · source ↗

    • dct
    • acceptability
    • blood
    • self-collection
    • capillary
    • pediatric
    • liquid
    • validation
  4. 2024

    Plasma separation cards showed high specificity (99.3%) but moderate sensitivity (87.5%) against EDTA-plasma for HIV-1 viral load monitoring at a 1,000 copies/ml threshold in a South African cohort. Despite favourable usability among healthcare workers, laboratory-reported technical difficulties must be addressed to enable reliable decentralised scale-up in resource-constrained settings.

    Assessment of the performance of the plasma separation card for HIV-1 viral load monitoring in South AfricaChimoyi et al., Journal of clinical microbiology · source ↗

    • blood
    • dried
    • capillary
    • validation
    • venous-agreement
    • acceptability
    • pediatric
    • serology
  5. 2023

    VAMS microsampling found no significant difference from venous blood for tacrolimus in paediatric patients; samples were stable for 14 days, with average difference between paired at-home samples of 0.12 ± 0.94 ng/mL.

    VAMS and parallel-reaction-monitoring mass spectrometry for tacrolimus trough measurements at home in paediatric heart-transplant patientsZhao et al., J. Mass Spectrometry and Advances in the Clinical Lab · source ↗

    • vams
    • venous-agreement
    • acceptability
    • blood
    • dbs
    • tdm
    • self-collection
    • dried
    • capillary
    • pediatric
    • immunosuppressants
    • validation